ContinueWe use cookies to give you the best experience on our website. You can browse the site with Carestream disabled cookies; However, your visit will be significantly improved if cookies will be active. If you continue without changing your settings, you agree to receive all cookies. However, you can change the setting of cookies at any time.

If you want to learn more (this is the link to our privacy policy, which has detailed information on how we use cookies and how to change their settings).

Photographic Processing Protocol for Autoradiography Emulsion, Type NTB

Use all processing solutions at a temperature of 15°C (59°F) for the times recommended in this table.

Use no agitation. 

Step  Recommended Solution  Time 
1. Develop KODAK DEKTOL Developer (1:1)
or Developer D-19 Substitute* (1:1)
(1:1 - One-to-one dilution with water)
2 minutes
4 minutes
2. Stop* Distilled Water 10 seconds
3. Fix** KODAK Fixer 5 minutes
4. Wash Distilled Water 5 minutes
5. Dry Slowly in a dust-free atmosphere  
*Do not use acid stop baths since they could result in the formation of microscopic bubbles in the emulsion layer.


**Do not use rapid fixers (which contain ammonium) since they could result in loss of developed silver grains.

Processing Chemicals

KODAK DEKTOL Developer and KODAK Fixer are Eastman Kodak Company products available through local photographic dealers.

Product List 

Cat. No.  Description  Unit Size 
889 5666 Autoradiography Emulsion Type NTB 118 mL (4 oz.)

Ordering Procedures

Autoradiography Emulsion, Type NTB orders can be accepted by fax, mail, or phone. Domestic orders (USA) may be placed as follows:
E-mail: molecular-orders@carestream.com
Phone: 877-747-4357
Fax; 1-800-879-4979
Outside the US: Phone: +1-203-786-5644
Outside the US Fax: + 1-203-786-5698

In the U.S., emulsions are shipped Monday through noon on Thursday by airfreight directly to the customer.

* D19 Developer Substitute is available from:
Photographers Formulary
Candon, MT 59826
Telephone: 1-406-754-2981